- 8661722 reads
Calculation of Results
FORMULA: ppm sulfite as SO3 = ml of potassium iodide-iodate (RXSOL-62-5503-006) x 500 / ml SAMPLE
Using a 50 ml sample, sulfite, in parts per million as SO3 is equal to the ml of RXSOL-62-5503-006 required multiplied by 10.
Example:
A 50-mL sample was titrated with 0.0125 N titrant and 6 mL of titrant was used at the endpoint. The concentration is 6 mL × 10 = 60 mg/L sulfite SO3 2–
Sample volumes and multipliers
Range (mg/L) | Sample volume (mL) | Titrant—potassium iodide-iodate | Multiplier |
0–100 | 50 | 0.0125 N | 10 |
40–200 | 25 | 0.0125 N | 20 |
100–500 | 10 | 0.0125 N | 50 |
More than 500 | 05 | 0.0125 N |
100 |
RX MARINE INTERNATIONAL is most popular chemicals source in INDIA, Keeping ready stock at 5 LOCATION which includes Mumbai, Kandla - Gandhidham, Vizag - Visakhapatnam, Ennore - Chennai, Kolkata more then 3000 types of chemicals. Please click here to get PRODUCT Categories details http://rxmarine.com/Products
Mail Us for any PRODUCT related information and purchase inquiry : rxmarinechemical@gmail.com ( This mail is for customer / Buyer only ) And if you wants to add your hand as a supplier, And if you have volume manufacturing capacity of chemicals, pl. drop your mail at rxmarinechemical@gmail.com ( This mail is for supplier only )